SBS issues notice for some CDOM fluorometers

Sea-Bird Scientific issued a notification that some CDOM fluorometers have been reported to measure biased values.  The bias has been traced back to three distinct root causes:

  1. Incorrect primary CDOM standard
  2. In situ bias
  3. Out-of-tolerance UV LED

For more information, please see their notification.